Transcript
Programming Basics Data Output Format ASCII Data Output Format This section describes the ASCII data output format, and the elements of the data. • “Time Stamp” • “Data Format” • “Data Elements” Time Stamp The B1500 can record the time when the measurement is started, and sends the time data (Time). This function is enabled by the TSC command. The time data will be sent just before the measurement data. For example, in the staircase sweep measurements, the data will be as shown below. Block1 [,Block2] . . . . <terminator> where, BlockN (N: integer) = Time1,Data1 [,Time2,Data2] ... [,Source_data], then TimeN (N: integer) is the time from timer reset to the start of DataN measurement. The timer count is cleared/reset by the TSR command (Time=0). The time stamp function is not available for the following measurements. • High speed spot measurement • Quasi-pulsed spot measurement (MM9) • Linear search measurement (MM14) • Binary search measurement (MM15) 1-26 Agilent B1500A/B1505A Programming Guide, Edition 11