Programming Examples - 4
Transient System Model
Figure 4-1 is a model of the transient system. The figure shows the transient modes and the source of the
data that generates each mode.
When a trigger is received in step or pulse modes, the triggered functions are set from their IMMediate to
their TRIGgered value. In Step mode, the triggered value becomes the immediate value. In Pulse mode,
the functions return to their immediate value during the low portion of the pulse. If there are no further
pulses, the immediate value remains in effect. In List mode, the functions return to their immediate value
at the completion of the list.
You can mix FIXed, STEP, PULSe, and LIST modes among most functions. When a trigger is received,
each function will react in a manner defined by its mode. However, this is subject to the following limitation
to ensure the proper output voltage in all cases:
NOTE:
The ac voltage, waveform shape, and voltage slew functions cannot be set to Step or
Pulse mode if one of them is set to List mode.
IMMediate level
Triggers ignored,
FIXED mode
output always set to
immediate command levels.
TRIGered level
At trigger, the triggered
IMMediate level
level becomes the new
STEP mode
immediate level.
TRIGered level
IMMediate level
At trigger, the triggered
PULSE mode
level is active during the
pulse width portion of the
pulse waveform
width
period
IMMediate level
At trigger, the list
LIST mode
When list completes, output
Step 2
returns to immediate level.
Step 0
Step 1
Trigger
Applied
List
Complete
Figure 4-1. Model of Transient System