Maintenance Symptom Symptom Number Four Erratic temperature display and outputs. • Target and measurement conditions unknown. Actual temperature variations may sensed. Possible Cause / Corrective Action 1. Check for recurrent interruptions in sight path e.g. bursts of smoke or steam, moving equipment. 2. Use air purge to clear some obstructions, If using air purge check air flow and operation 3. Vary the Response Time to see if the symptom changes. 4. Use Peak Picker signal conditioning with suitable decay rate. 5. Use Track and Hold signal conditioning. Symptom Number Five Erratic display and outputs. • Target temperature and measurement conditions known to be stable 1. Check Sensor cable connections. Also, check the signal cable shield connection at the Grounded Strain Relief Fitting. 2. Check for proper grounding of all system components. 3. Check cable routing. Signal cables must not run in the same conduit as noisy power lines or power lines with transients. 4. Check ambient temperature of Sensor Add water cooling and/or heat shielding to Sensor, if necessary. No Invalid Measurement Condition Alarm 1. Check iALM function under Engineering Menu. Cannot adjust or select functions at the Sensor rear panel. 1. Check Panel Lock function in Engineering Menu. RS-485 Communications problems 1. Check Communications Lock setting in COMMS Menu. Selections are Read Only and Read / Write. 2. Check Baud Rate setting in COMMS Menu. 3. Check Address setting in COMMS Menu. Laser will not energize 1. Laser will not energize when CAL function is operating or initiated Peak Picker signal conditioning not functioning 1. Check if the Auto Peak Picker Reset is On. If ON, evaluate the Reset Below Temperature settings. Peak Picker may be Auto reset with Reset Below temperature too low Peak Picker Delay not operating 1. Peak Picker Delay only operates after a Peak Picker Reset or after the first temperature indication displayed after an Invalid (INV) condition or below Sensor temperature range (LOW) condition. 2. Delay timing starts immediately after the reset or the Invalid or LOW condition. Can not adjust AOUT ZERO or FULL Scale 136 1. Check both ZERO and FULL settings. Minimum 10°F span is required (or equivalent °C span values) Rev. L4 12/2012 Modline 5