Sensor Installation
Dimensions are in inches and (millimeters). Do not scale.
Use 5/16 inch or M8 mounting Bolts (3 Places).
3.9.8 ESA – Extension Sleeve Adapter Description and Dimensions
This is a special accessory for Sensors with the Dirty Window Detector (DWD) option. The ESA serves
as protection sleeve over the DWD optics and as the mechanical interface between the Sensor and
other Modline 5 accessories. The ESA is used in the following combinations to provide air purging or
create a dead air space zone in front of the Sensor optics to help keep them clean. The combinations
also provide shielding from stray radiation emanating from high temperature targets. The APA air
purge, shown in combinations one and two, provides the best performance. The use of an APA is
required for large targets over 1000 °C.
1.
ESA and APA air purge with clean air (air purging and shielding).
2.
ESA and APA air purge without air (dead air space zone and shielding).
3.
ESA with MFL Flange (dead air space zone and shielding)
4.
ESA with MFL and AA-3 air purge accessory (air purging and shielding).
See Section 3.8 DWD Installation Requirements for DWD and Section 3.12 Other Accessories (Sight
Tube, Window, WJ-5 Water Cooling), page 59 for AA-3 references.
Figure 37
The ESA is also used with the MFL Flange when as an interface with the Right Angle Mounting Bracket.
Material: Clear Anodized Aluminum
Weight: Approximately 1 Pound (0.5 kg)
Mounting and Assembly Supplied Parts List:
• No hardware is supplied with the ESA.
• Assembly screws are supplied with the APA and MFL.
48
Rev. L4 12/2012
Modline 5