Download Manual - StanTronic Instruments
Transcript
9.1 Output Output terminals Terminal block with M4 screws (rear panel) AC receptacle (universal type, front panel) Note: Use AC receptacle for AC (0 V AC to 250 V AC). When DC is included, use screw terminal block on rear panel. Rated output 100 V range: 100 Vrms (AC) / 100 V (DC) 200 V range: 200 Vrms (AC) / 200 V (DC) Output voltage settings (When signal source mode is internal mode or internal + external mode, no load) Note: The AC settings (peak value)+ DC setting that can be set are within the voltage setting limit range. AC 100 V range: 0.0 to 140.0 Vrms 200 V range: 0.0 to 280.0 Vrms Setting resolution: 0.1 Vrms Setting accuracy: ±(0.5% of set value +0.6 Vrms): 100 V range ±(0.5% of set value +1.2 Vrms): 200 V range (AC mode, 50 Hz, 23 ±5°C) DC 100 V range: 200.0 V to +200.0 V 200 V range: 400.0 V to +400.0 V Setting resolution: 0.1 V Setting accuracy: ±( 0.5 % of set value + 0.8 V): 100 V range ±( 0.5 % of set value + 1.6 V): 200 V range (AC+DC mode, 0 V AC setting, 23 ±5°C) Maximum output current 100 V range: 10 Arms (AC) / 10 A (DC) 200 V range: 5 Arms (AC) / 5 A (DC) Note 1: For the AC 100 V input system, the limit on the maximum output power may cause a reduction in the maximum output current. Note 2: At or above the rated output voltage, the limit on the maximum output power reduces the maximum output current. Note 3: When output frequency is 40 Hz or lower, or it is 400 Hz or higher, the maximum output current may decrease. Note 4: The RMS value for AC + DC is within the maximum output current. 9-3 APS-1102 Programmable AC/DC Power Source