Download Manual - StanTronic Instruments
Transcript
9.11 External Control I/O 9.11 External Control I/O External control operation mode Enabled or disabled (status output is continually output) Control input Input level High level: +4.0 V or above, Low level: +1.0 V or below Nondestructive maximum input: +10 V/5 V Input impedance +5 V pull-up with 47 k Control item Output off: Output off at fall Output on: Output on at fall Start sequence: Start at fall Stop sequence: Stop at fall Hold input: Hold at fall Branch inputs 0, 1: Start branch at fall Status output Output level 0/+5 V (open) Output impedance 100 Ω Status items Power source on/off status: 0-off, 1-on Output on/off status: 0-off, 1-on Limiter operation: 0- Not operating, 1- Operating Software busy: 0- Normal, 1- Busy Sequence operation step synchronized outputs 0 and 1 Terminals D-sub 25-pin multi-connector (rear panel) 9.12 USB Interface USB interface is provided to enable control from an external computer. Interface standard: USBTMC 9.13 Power Input Input voltage range: 100 V AC to 230 V AC ±10% (not exceeding 250 V) Overvoltage Category II Frequency range: 50 Hz or 60 Hz ±2 Hz (single phase) Power consumption: 1.4 kVA or less Power factor: 0.95 or above (during 100 V AC input) 0.90 or above (during 200 V AC input) Protective ground impedance: 0.1 Ω or less (not including the power cord set) 9-13 APS-1102 Programmable AC/DC Power Source