Download CerePlex Direct - Blackrock Microsystems

Transcript
Blackrock CerePlex Direct - Instruction for Use
Digital Out
This window will allow the user to set the properties of the digital outputs transmitted to other equipment connected
to the Host PC.
Figure 5-9 - Digital Output Properties

Label: A user-defined name for the specific channel.

Properties: The user has the option to Monitor Electrodes, which will send a TTL pulse from that specific
Digital Output every time a spike is detected on a specific channel or unit of a channel. As a second option,
the user can select Timed to output a user specified frequency range with a specified duty cycle and offset.
The low and high samplings show the number of clock cycles the square wave is set to high or low. The
offset unit is clock cycles and can offset the trigger of one channel from the other output channels. Each clock
cycle is 1/30,000 seconds long.

Track Most Recently Selected: When this box is checked the channel source will change based on whichever
channel is selected from Spike Panel, Single Neural Channel, or Raster Plot.

Source: Select a source channel to monitor.

Monitor Units: The user can select which units to monitor from the source channel through the digital output
ports. The user can monitor unclassified units and/or classified units defined online.
LB-0462
3.00
Page 21 of 53