Nicolet Evolution 500 Local Control Software Manual
* LAMPS *
TUNGSTEN
HOURS
ENERGY
D2
HOURS
ENERGY
Hg
HOURS
ENERGY
: ON
: 987
: 99%
: STRIKING
: 234
: 96%
: OFF
:2
: 0%
RESET HOURS – Displays a menu to choose whether the
tungsten, deuterium or mercury lamp is to be reset. When
selected resets hours and measures energy at the
appropriate wavelength. ALLOW THE LAMP AT LEAST 10
MINUTES TO WARM UP BEFORE RESETTING ITS
HOURS.
ENERGY - Displays a menu to choose whether the energy of
the tungsten, deuterium or mercury lamp is to be measured.
ALLOW THE LAMP AT LEAST 10 MINUTES TO WARM UP
BEFORE MEASURING ITS ENERGY. CLEAR BOTH
SAMPLE AND REFERENCE BEAMS BEFORE
MEASURING LAMP ENERGIES.
SWITCH Hg – Switches the Mercury lamp ON or OFF
depending on its current status.
THE MERCURY LAMP SHOULD BE TURNED OFF WHEN
NOT IN USE.
SWITCH D2 -Turns the Deuterium lamp ON or OFF
depending on its current status.
19
Issue 1(August 2003)