Download SCAMAX - InoTec

Transcript
4 COMMUNICATION
If communication problems arise between the scanner and the scan software (usually caused by some
incompatibility of the standard ISIS or TWAIN) these can be analysed using a trace file. The trace file can
be created at various points within the communication path (refer diagram on next page).
To determine the PC configuration a status file can be generated by the scan PC. To do this start the
DPUSTAT32.exe program found on the driver CD in the folder Utils.
DPUSTAT32.exe
SCAMAX®H04
Twain-Driver
(InoTec_M03.ds)
Twain-Application
ISIS-Driver
(Ino_M03.pxn)
ISIS-Application
Aspi
DPU-Driver
(JuK_Spti.di2)
API
Software
Hardware
DPU-Scan
Adaptec 29160
Service
RS 232
SCSI-Board
(128-512 MB)
Service
Controller
RS 232
Command
SCSI 160
RS 232
I/O
PC
Scanner
Display
RS 232
Mechanic/
Optic
Endorser
Communication Pathes
I²C
Camera
Front
Video
I²C
Camera
Back
Video
…\PIXTRAN\Ino_M03.cfg
[DEBUG]
LogFile=C:\isis.log
…\DpuDrvs\DPU.INI
DiagOutput=1
TraceMessage=1
[TWDPU.DS]
LogEnabled=1
TraceForward=1
LogFileName=C:\twain.log
…\DpuDrvs\DPU.INI
[DPUINIT]
LogEnabled=1
LogFileName=C:\dpu.log
LogDbgOut=1
ScanDiag.exe
>>Trace
Analyze Tools
The actual fault analysis is done by InoTec. Please email the trace and/or status file generated to
[email protected] .
Technical Manual 1/2012
57