Figure 1.3: Position of lasers. Invisible rack detection lasers are located within
the rectangle area. The 2D code reader (visible) is located within the open
circle.
1.4 Secure installation
This section describes the requirements your site must meet for
safe installation and operation of your autoMACS Pro Separator.
Read the instructions in this section and ensure that your site is
properly prepared before you connect the instrument to its power
source. When planning your site layout and equipment locations,
keep in mind the precautions described in this section to help avoid
instrument failures and reduce the possibility of environmentally
caused shutdowns.
1.4.1 Mounting accessories
Do not place the autoMACS Pro Separator on an unstable table,
cart, stand, tripod, or bracket. As a consequence, the instrument
might fall down. This may cause serious bodily harm and/or serious
damage to the instrument. Use only on a table, cart, stand, tripod,
or bracket recommended by Miltenyi Biotec or sold with the
instrument. Do not place the autoMACS Pro Separator within a
built-in apparatus or a confined space such as a shelf rack unless
the apparatus has been specifically designed to accommodate
the instrument, proper ventilation is provided, and the mounting
instructions for the instrument have been followed.
1 Important information
17