Download Vectorbeam Series
Transcript
Page 171 of these measurements (uncorrected) and the average of these measurements corrected for the edge sharpness (corrected). The parameters FILT, POINT and LINESCANS must be chosen to be large enough so that noise on the signal does not make the measurement inaccurate. 17.6.2.1. Measure diameter using metal calibration mark Measuring the diameter on the calibration mark is relatively quick, can be done automatically and can be done on any holder. It is limited to spot sizes above about 30 nm. 1. Move stage to the 10 µm octagonal calibration mark. 2. With a 16-bit pattern generator in order to be able to scan the beam in small steps to measure the smallest spot sizes more accurately the pattern generator resolution should be set to a small value. Set the maximum fieldsize to e.g. 0.16384 mm: VB_OPER> FLD 0.16384 For the 18-bit and 20-bit pattern generators this is not necessary. 3. Switch to SEM mode and graph mode. 4. Adjust gain and backoff so that the signal from the mark is a little below the maximum level and the signal from the background is a little above the zero level. 5. Adjust the focus and stigmation. 6. Type : VB_OPER>QDISPLAY DIA FM / ALL / LINELENGTH=2 / EDGE=5 / FILT=16 / POINT=8 / LINESCANS=1 17.6.2.2. Measure diameter using knife edge Measuring the diameter using a knife edge is a relatively long procedure, must be done manually and usually only one holder has a knife edge loaded. It does however allow spot sizes below 10 nm to be measured. A fresh knife edge can be made by cleaving a piece of GaAs wafer so as to give a smooth face normal to the wafer surface. Care must be taken to ensure the face is smooth, has no damage to the top edge and has little debris on the edge. The piece must be small enough to mount on the knife edge holder. The edge is nominally in the same plane as the substrate surface in the holder but because the knife edge is not front face referenced, the edge may be a few hundred microns above or below the plane. 1. Load holder containing the knife edge on the stage. 2. With a 16-bit pattern generator in order to be able to scan the beam in small steps to measure the smallest spot sizes more accurately the pattern generator resolution should be set to a small value. Set the maximum fieldsize to e.g. 0.16384 mm : VB_OPER> FLD 0.16384 3. For the 18-bit and 20-bit pattern generators this is not necessary. Part Number:878275 Vectorbeam Operator Manual