Commands Listed in Alphabetical Order
Commands that affect BUSY? response
Examples
Operation
Command
Single sequence
acquisition
ACQuire:STATE ON or
ACQuire:STATE RUN or
ACQuire:STATE1
(when ACQuire:STOPAfter is set to SEQuence)
Hard copy operation
HARDCopy STArt
Calibration step
Refer to the optional oscilloscope Service Manual.
BUSY? might return :BUSY 1 indicating that the oscilloscope is currently busy.
*CAL? (Query Only)
Performs an internal self-calibration and returns the oscilloscope calibration status.
NOTE. Disconnect or otherwise remove all input signals prior to starting
self-calibration. The self-calibration can take several minutes to complete.
No other commands are executed until calibration is complete.
Group
Calibration and Diagnostic
Syntax
*CAL?
Returns
<NR1> = 1 indicates the calibration did not complete successfully.
<NR1> = 0 indicates the calibration completed without errors.
Examples
*CAL? starts the internal signal path calibration and might return 0 to indicate that
the calibration was successful.
CALibrate:FACtory:STATus? (Query Only)
Returns the factory calibration status value saved in nonvolatile memory.
2-144
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STATus?
MSO3000 and DPO3000 Series Programmer Manual