Commands Listed in Alphabetical Order Commands that affect BUSY? response Examples Operation Command Single sequence acquisition ACQuire:STATE ON or ACQuire:STATE RUN or ACQuire:STATE1 (when ACQuire:STOPAfter is set to SEQuence) Hard copy operation HARDCopy STArt Calibration step Refer to the optional oscilloscope Service Manual. BUSY? might return :BUSY 1 indicating that the oscilloscope is currently busy. *CAL? (Query Only) Performs an internal self-calibration and returns the oscilloscope calibration status. NOTE. Disconnect or otherwise remove all input signals prior to starting self-calibration. The self-calibration can take several minutes to complete. No other commands are executed until calibration is complete. Group Calibration and Diagnostic Syntax *CAL? Returns <NR1> = 1 indicates the calibration did not complete successfully. <NR1> = 0 indicates the calibration completed without errors. Examples *CAL? starts the internal signal path calibration and might return 0 to indicate that the calibration was successful. CALibrate:FACtory:STATus? (Query Only) Returns the factory calibration status value saved in nonvolatile memory. 2-144 Group Calibration and Diagnostic Syntax CALibrate:FACtory:STATus? MSO3000 and DPO3000 Series Programmer Manual