WFMOutpre:BIT_Nr <NR1>
WFMOutpre:BIT_Nr?
DATa:SOUrce,
DATa:WIDth,
WFMOutpre:BYT_Nr
Arguments
Examples
<NR1> is the number of bits per data point and can be 8 or 16.
WFMOUTPRE:BIT_NR 16 sets the number of bits per waveform point to 16 for
outgoing waveforms.
WFMOUTPRE:BIT_NR? might return :WFMOUTPRE:BIT_NR 8 indicating that
outgoing waveforms use 8 bits per waveform point.
WFMOutpre:BN_Fmt
This command specifies the format of the binary data for outgoing waveforms
when WFMOutpre:ENCdg is set to BINary. The format can be RI (signed
integer) or RP (positive integer) for analog channels. Changing the value of
WFMOutpre:BN_Fmt also changes the value of DATa:ENCdg. The waveform is
specified by the DATa:SOUrce command.
(See page 2-78, Waveform Transfer Command Group.)
(See page D-1, Waveform Transfer (WFMOutpre and CURVe Query) Examples.)
Group
Waveform Transfer
Syntax
WFMOutpre:BN_Fmt {RI|RP}
WFMOutpre:BN_Fmt?
Related Commands
DATa:ENCdg,
DATa:SOUrce
Arguments
RI specifies signed integer data point representation.
RP specifies positive integer data point representation.
Examples
2-594
WFMOUTPRE:BN_FMT RP specifies that outgoing waveform data will be in positive
integer format.
MSO3000 and DPO3000 Series Programmer Manual
Related documents
Tektronix DPO4000 User's Manual
Programmer Manual
TRIGger:A:BUS:B
Oscilloscope Programmer`s Reference
Connect-Tek USB-2 Specifications
Datasheet - Tektronix
Data Sheet
MSO3000 and DPO3000 Series Oscilloscopes User Manual
InPEduS+Avionics: User Manual
DLM4000 Series Mixed Signal Oscilloscope - Electro
DLM2000 Series Digital Oscilloscope/Mixed Signal
Bench Products Catalog