Commands Listed in Alphabetical Order
Conditions
This command requires a DPO3PWR application module.
Group
Power
Syntax
POWer:REFLevel:ABSolute:HIGH <NR3>; Ranges={D,-1e6,+1E6}
POWer:REFLevel:ABSolute:HIGH?
Arguments
Examples
<NR3> is the absolute high value in volts. Default value is 0.0E+0
POWER:REFLEVEL:ABSOLUTE:HIGH 2 sets the absolute high value to 2.000.
POWER:REFLEVEL:ABSOLUTE:HIGH? might return
POWER:REFLEVEL:ABSOLUTE:HIGH 2 indicating that the absolute high
value is set to 2 V.
POWer:REFLevel:ABSolute:LOW
Sets or returns the low reference level for power measurements.
Conditions
This command requires a DPO3PWR application module.
Group
Power
Syntax
POWer:REFLevel:ABSolute:LOW <NR3>; Ranges={D,-1e6,+1E6}
POWer:REFLevel:ABSolute:LOW?
Arguments
Examples
<NR3> is the absolute low value in volts. Default value is 0.0E+0.
POWER:REFLEVEL:ABSOLUTE:LOW 1.0 sets the absolute high value to 1.0000.
POWER:REFLEVEL:ABSOLUTE:LOW? might return
POWER:REFLEVEL:ABSOLUTE:LOW 1.0000 indicating that the absolute low
value is set to 1 V.
POWer:REFLevel:ABSolute:MID<1-3>
Sets or returns the mid reference level for measurements. MID3 is specific to
the power application.
MSO3000 and DPO3000 Series Programmer Manual
2-327
Related documents
Tektronix DPO4000 User's Manual
Programmer Manual
TRIGger:A:BUS:B
Oscilloscope Programmer`s Reference
Connect-Tek USB-2 Specifications
Datasheet - Tektronix
Data Sheet
MSO3000 and DPO3000 Series Oscilloscopes User Manual
InPEduS+Avionics: User Manual
DLM4000 Series Mixed Signal Oscilloscope - Electro
DLM2000 Series Digital Oscilloscope/Mixed Signal
Bench Products Catalog