POWER:HARMONICS:IEC:INPUTPOWER 600 sets the class D input power to
600 W.
POWER:HARMONICS:IEC:INPUTPOWER? might return :POWER:HARMONICS:
IEC:INPUTPOWER 100 indicating that the class D input power is set to 100 W.
POWer:HARMonics:IEC:LINEFREQuency
Sets or returns the line frequency for the IEC standard.
Conditions
This command requires a DPO3PWR application module.
Group
Power
Syntax
POWer:HARMonics:IEC:LINEFREQuency <NR3>
POWer:HARMonics:IEC:LINEFREQuency?
Arguments
Examples
<NR3> is the line frequency. The valid values are 50 and 60.
POWER:HARMONICS:IEC:LINEFREQUENCY 50 sets the line frequency to 50 Hz.
POWER:HARMONICS:IEC:LINEFREQUENCY? might return :POWER:
HARMONICS:IEC:LINEFREQUENCY 60 indicating that the line frequency value
is set to 60 Hz.
POWer:HARMonics:IEC:OBSPERiod
Sets or returns the IEC observation period.
Conditions
This command requires a DPO3PWR application module.
Group
Power
Syntax
POWer:HARMonics:IEC:OBSPERiod <NR3>
POWer:HARMonics:IEC:OBSPERiod?
Arguments
<NR3> is the IEC observation period. Valid values range from 0.2 to 10 s.
MSO3000 and DPO3000 Series Programmer Manual