Commands Listed in Alphabetical Order Examples POWER:HARMONICS:IEC:INPUTPOWER 600 sets the class D input power to 600 W. POWER:HARMONICS:IEC:INPUTPOWER? might return :POWER:HARMONICS: IEC:INPUTPOWER 100 indicating that the class D input power is set to 100 W. POWer:HARMonics:IEC:LINEFREQuency Sets or returns the line frequency for the IEC standard. Conditions This command requires a DPO3PWR application module. Group Power Syntax POWer:HARMonics:IEC:LINEFREQuency <NR3> POWer:HARMonics:IEC:LINEFREQuency? Arguments Examples <NR3> is the line frequency. The valid values are 50 and 60. POWER:HARMONICS:IEC:LINEFREQUENCY 50 sets the line frequency to 50 Hz. POWER:HARMONICS:IEC:LINEFREQUENCY? might return :POWER: HARMONICS:IEC:LINEFREQUENCY 60 indicating that the line frequency value is set to 60 Hz. POWer:HARMonics:IEC:OBSPERiod Sets or returns the IEC observation period. Conditions This command requires a DPO3PWR application module. Group Power Syntax POWer:HARMonics:IEC:OBSPERiod <NR3> POWer:HARMonics:IEC:OBSPERiod? Arguments <NR3> is the IEC observation period. Valid values range from 0.2 to 10 s. MSO3000 and DPO3000 Series Programmer Manual 2-301