6. Visually average the two traces; imagine a trace directly between the two traces. (If both
traces are not completely visible, change the scale as required.)
Press 4
5 rotate the RPG knob, and record the maximum \averaged" power deviation
from 10 kHz to 5 MHz on the performance test record.
7. Press 4
5 DEFINE TRACE TRACE: DATA .
Marker
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Display
8. Change the connection as shown in Figure 2-32.
Figure 2-32. Magnitude Ratio Frequency Response Test Setup-4
9. Set the E5100A as follows:
Control Settings
Preset
Power= 04 dBm
Input-R Attenuator: 0 dB
Input-A Attenuator: 0 dB
Key Strokes
4Preset5
4Sweep5
4System5
NNNNNNNNNNNNNNNNN
POWER
1
MORE ATTENUATOR PORT : R 0 dB
A 0 dB
2-64 Performance Tests for E5100A/B Option 100/200/300/400/600
NNNNNNNNNNNN
4
NNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNN
NNNNN NNNNNNNNNNNNNN
10. Repeat the steps 3 through 6.
11. Repeat test for B/R (Opt.300/400).
12. Repeat test for B/C (Opt.400).