Download HP E5100A,B Service
Transcript
6. Visually average the two traces; imagine a trace directly between the two traces. (If both traces are not completely visible, change the scale as required.) Press 4 5 rotate the RPG knob, and record the maximum \averaged" power deviation from 10 kHz to 5 MHz on the performance test record. 7. Press 4 5 DEFINE TRACE TRACE: DATA . Marker NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN Display 8. Change the connection as shown in Figure 2-32. Figure 2-32. Magnitude Ratio Frequency Response Test Setup-4 9. Set the E5100A as follows: Control Settings Preset Power= 04 dBm Input-R Attenuator: 0 dB Input-A Attenuator: 0 dB Key Strokes 4Preset5 4Sweep5 4System5 NNNNNNNNNNNNNNNNN POWER 1 MORE ATTENUATOR PORT : R 0 dB A 0 dB 2-64 Performance Tests for E5100A/B Option 100/200/300/400/600 NNNNNNNNNNNN 4 NNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNN NNNNN NNNNNNNNNNNNNN 10. Repeat the steps 3 through 6. 11. Repeat test for B/R (Opt.300/400). 12. Repeat test for B/C (Opt.400). 052 4