Check A27 Synthesizer Output One of the output signals from the A27 Synthesizer is INT REF signal on the rear panel. The input signal to A27 is the external reference signal from the EXT REF connector. If the output signal and the E5100A/B operation using the EXT REF input signal are good, A27 is probably good. Perform the following procedures sequentially to verify all the signals listed above and to verify the E5100A/B operation when the EXT REF signal is used. 1. Check the INT REF Signal a. Connect the equipment as shown in Figure 9-1. Figure 9-1. INT REF Test Setup b. Initialize the spectrum analyzer. Then set the controls as follows: Controls Settings Center Frequency 10 MHz Span 15 MHz Reference Level 10 dBm 5 to move the marker to the peak of the INT c. On the spectrum analyzer, press 4 REF signal. d. Check that the frequency is approximately 10 MHz and the level is +2 dBm 6 4 dB. The INT REF signal should be as shown in Figure 9-2. If the INT REF signal is good, continue with 3. Check the FRAC N OSC Signal. If the INT REF signal is bad, inspect the cable and connections between the INT REF connector and A27J10. If the cable and connections are good, replace A27. PEAK SEARCH 9-6 Source Group Troubleshooting