31. Without changing any measuring receiver settings, set the UUT to the next amplitude
listed in Table 5-14 for this frequency. Allow the measuring receiver reading to
stabilize and record the reading as Prel. (Note that Prel will be a negative number in
dB).
32. Calculate the UUT output Pout = P-37 + Prel. Check that the value of Pout is within
the tolerance shown in part A of Table 5-14.
33. Repeat steps 31 and 32 for each amplitude listed in part A of Table 5-14 for this
frequency.
34. Return the UUT output to -37dBm, set the UUT frequency to the next frequency
listed in part A of Table 5-14, and repeat steps 30 through 34.
35. Set the UUT to 100kHz at -47dBm. Repeat steps 30 through 35 for the points listed
in part B of Table 5-14 using the calculation Pout = P-47 + Prel in step 32.
36. Set the UUT to STBY.
Table 5-14. Level Accuracy Test (50•), Low Level Test Points
Part
Frequency
A
100 kHz
10 MHz
125 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
5-26
Reading
Pout Tolerance
-37 dBm
Reference
Reference
-66 dBm
Prel
±0.20 dBm
-85 dBm (P-85)
Prel
±0.50 dBm
-37 dBm
P-37
Reference
-66 dBm
Prel
±0.20 dBm
-85 dBm (P-85)
Prel
±0.50 dBm
-37 dBm
Reference
Reference
-66dBm
Prel
±0.20 dBm
-85 dBm (P-85)
Prel
±0.50 dBm
-37 dBm
P-37
Reference
-66dBm
Prel
±0.20 dBm
-85 dBm (P-85)
Prel
±0.50 dBm
-37 dBm
Reference
Reference
-66dBm
Prel
±0.50 dBm
-85 dBm (P-85)
Prel
±1.00 dBm
-37 dBm
P-37
Reference
-66dBm
Prel
±0.50 dBm
-85 dBm (P-85)
Prel
±1.00 dBm
-37 dBm
Reference
Reference
-66dBm
Prel
±0.50 dBm
-85 dBm (P-85)
Prel
±1.00 dBm
Amplitude