Remote Operation
Preparing the Instrument for Remote Operation
4A
Table 4A-1. GPIB Preferences
ead98f.bmp
Field
GPIB Address
Event Status Enable
Status Register Enable
Power On Status Clear
Preference
0 to 30
0 to 255 (May also be set using SCPI *ESE command.)
0 to 255 (May also be set using SCPI *SRE command.)
Enable, Disable
Switching to Remote Operation
When the Instrument is in local operation the instrument is fully programmable both from
the front panel and from the IEEE 488 bus. There is no physical switch for selecting
remote operation. Rather, when the Instrument receives a bus message it switches to
remote operation. If the message arrives while a change is being entered from the front
panel, the front panel entry is interrupted and then the bus message is executed. Once the
Instrument is set to remote operation all of the front panel (local) controls are locked out
(inoperable) with the exception of the L key and the Go to Local soft key at the bottom
of the display. Pressing the GO to Local soft key causes the Instrument to return to local
operation.
4A-5