R&S®UPP Maintenance and Instrument Interfaces Instrument Interfaces The number of controllers which can access an instrument is virtually unlimited in the network. In the instrument, a clear distinction is made between the individual controllers. This distinction is continued down to one controller at application level, i.e. two applications on a PC are regarded as two different controllers by an instrument. Fig. 9-4: Remote control via LAN with several controllers The controllers can lock and unlock the instrument to permit exclusive access. As a result, it is possible to control access from several controllers. In the program 'Measurement and Automation Control', this setting is made in the Template panel. 9.4.2.1 VXI-11 interface messages With Ethernet connections, the interface messages are referred to as low-level control messages. They can be used to emulate the interface messages of the IEC/IEEE bus. User Manual 1411.1055.32 ─ 03 805