Download Agilent Technologies 2000 X-Series Technical data

Transcript
:TRIGger Commands
29
:TRIGger:LEVel:ASETup
(see page 790)
Command Syntax
:TRIGger:LEVel:ASETup
The :TRIGger:LEVel:ASETup command automatically sets the trigger levels
of all displayed analog channels to their waveforms' 50% values.
If AC coupling is used, the trigger levels are set to 0 V.
When High and Low (dual) trigger levels are used (as with Rise/Fall Time
and Runt triggers, for example), this command has no effect.
See Also
• ":TRIGger[:EDGE]:LEVel" on page 603
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
595